Degreasing and descaling agent for the routine service cleaning of all bench-top autoclaves
To be diluted
Concentrated formula
• Drain the autoclave reservoir.
• Mix 20ml (2 caps full) of System4 concentrate to 1 litre of demineralised water. In cases of particularly heavy soiling use double strength i.e. 40ml (4 caps full) to 1 litre of demineralised water.
• Refill reservoir with the cleaning solution.
• Run autoclave for one cycle – following cycle completion, drain reservoir of all remaining System4 solution.
• Refill reservoir with demineralised water and run autoclave for one further cycle to rinse.
• Following rinse cycle drain reservoir and wipe interior chamber dry.
• Refill reservoir ready for use.